HPLC
8 products
Browse 8 HPLC models from Labomiz Scientific, organised into 1 product families. Open any model for full specifications, download its catalog, or request a quotation. Tick up to four products to compare them side by side.
All HPLC models
High performance liquid chromatography MHPLC-1A
| Pump specifications | Double plunger reciprocating pump Isocratic Pump (1 set) Flow range 0.001 mL / min ̴ 10.000 mL / min Setting step size 0.001 mL / min Flow accuracy ≤ ± 0.3 % ( 1.00 ml / min, H2O , R.T. , 8.5 MPa ) Flow stability RSD ≤ 0.1 % ( 1.00 ml / min, H2O , R.T , 8.5 MPa ) Maximum pressure 42 Mpa Display pressure error ≤ ± 3.0 % Pressure fluctuation ≤ 1.0 % Power dissipation 80 W |
| Detector specifications | Wavelength repeatability ± 0.1 nm Wavelength accuracy ≤ ± 1.0 nm Spectrum band width 8 nm Response time 0.0 secs to 4.9 secs Linearity range 1.8 AU ( 5 % ) Baseline noise ≤ ± 1 × 10 - 5 AU Wavelength range 190 nm ~ 700 nm Baseline drift ≤ 2.0 × 10-4 AU / h Minimum detectable concentration 4.0 × 10-9 g / mL ( Naphthalene in Methanol Solution ) Detector cell Optical Path - 10 mm Volume - 10 μl Light source Deuterium Lamp Power dissipation 100 W Power supply AC 110 V / 220 V ± 10 %, 50 Hz / 60 Hz |
High performance liquid chromatography MHPLC-1B
| Pump specifications | Double plunger reciprocating pump Binary high pressure gradient pump(2 set) Flow range 0.001 mL / min ̴ 10.000 mL / min Setting step size 0.001 mL / min Flow accuracy ≤ ± 0.3 % ( 1.00 ml / min, H2O , R.T. , 8.5 MPa ) Flow stability RSD ≤ 0.1 % ( 1.00 ml / min, H2O , R.T , 8.5 MPa ) Maximum pressure 42 Mpa Display pressure error ≤ ± 3.0 % Pressure fluctuation ≤ 1.0 % Power dissipation 80 W |
| Detector specifications | Wavelength repeatability ± 0.1nm Wavelength accuracy ≤ ± 1.0 nm Spectrum band width 8 nm Response time 0.0 sec ̴ 4.9 sec Linearity range 1.8 AU ( 5 % ) Baseline noise ≤ ± 1×10 – 5 AU Wavelength range 190 nm - 700 nm Baseline drift ≤2.0 × 10 - 4 A U / h Minimum detectable concentration 4.0×10-9 g / mL (Naphthalene in Methanol Solution) Detector cell Optical Path—10 mm Volume—10 μL Light source Deuterium Lamp Power dissipation 100 W Power supply AC 110 V / 220 V ± 10 %, 50 Hz / 60 Hz |
High performance liquid chromatography MHPLC-2A
| Pump specifications | Pump Isocratic Pump(1 Set) Flow rate range 0.001 mL / min ̴ 10.000 mL / min Flow accuracy ≤ ± 0.2 % (1.00 ml / min, H2O, R.T., 8.5 MPa) Flow precision RSD ≤ 0.075 % (1.00 ml / min, H2O, R.T., 8.5 MPa) Maximum pressure 45 MPa Pressure accuracy ≤ ± 3.0 % Pressure pulsation ≤ 0.1 MPa |
| Detector specifications | Wavelength range 190 nm ~ 700 nm Wavelength repeatability ≤ 0.1 nm Wavelength accuracy ≤ ± 1.0 nm Band width 8 nm Response time 0.0 sec ̴ 5 sec Baseline noise ≤ ± 0.5 × 10 – 5 AU Drift ≤ 1.0 × 10 -4 AU / h Path length 10 mm (analytical cell) Light source Deuterium Lamp |
High performance liquid chromatography MHPLC-2B
| Pump specifications | Pump Binary high pressure gradient pump (2 Set) Flow Rate Range 0.001 mL / min ̴ 10.000 mL / min Flow Accuracy ≤ ± 0.2 % ( 1.00 ml / min, H2O, R.T., 8.5 Mpa ) Flow Precision RSD ≤ 0.075 % (1.00 ml / min, H2O, R.T., 8.5 Mpa ) Maximum Pressure 45 MPa Pressure accuracy ≤ ± 3.0 % Pressure pulsation ≤ 0.1 MPa |
| Detector specifications | Wavelength range 190 nm – 700 nm Wavelength repeatability ≤ 0.1 nm Wavelength accuracy ≤ ± 1.0 nm Band width 8 nm Response time 0.0 sec ̴ 5 sec Baseline noise ≤ ± 0.5 ×10-5 AU Drift ≤ 1.0 × 10-4 AU / h Path length 10 mm (analytical cell) Light source Deuterium Lamp Power supply AC 110 V / 220 V ± 10 %, 50 Hz / 60 Hz |
High performance liquid chromatography MHPLC-2C
| Pump specifications | Pump Quaternary low pressure gradient pump(1 Set) Flow rate range 0.001 mL / min ̴ 10.000 mL / min Flow accuracy ≤ ± 0.2 % (1.00 ml / min, H2O, R.T., 8.5 MPa) Flow precision RSD ≤ 0.075 % (1.00 ml / min, H2O ,R.T., 8.5 MPa) Maximum pressure 45 MPa Pressure accuracy ≤ ± 3.0 % Pressure pulsation ≤ 0.1 MPa |
| Detector specifications | Wavelength range 190 nm ̴ 700 nm Wavelength repeatability ≤ 0.1 nm Wavelength accuracy ≤ ± 1.0 nm Band width 8 nm Response time 0.0 sec ̴ 5 sec Baseline noise ≤ ± 0.5 × 10 - 5 AU Drift ≤ 1.0 × 10-4 AU / h Path length 10 mm (analytical cell) Light source Deuterium Lamp Power supply AC 110 V / 220 V ± 10 %, 50 Hz / 60 Hz |
High performance liquid chromatography MHPLC-3A
| Pump specifications | Pump Quaternary low pressure gradient pump(1 Set) Flow rate range 0.001 mL / min ̴ 10.000 mL / min Flow accuracy ≤ ± 0.2 % (1.00 ml / min, H2O, R.T., 8.5 MPa) Flow precision RSD ≤ 0.075 % (1.00 ml / min, H2O ,R.T., 8.5 MPa) Maximum pressure 45 MPa Pressure accuracy ≤ ± 3.0 % Pressure pulsation ≤ 0.1 MPa |
| Detector specifications | Wavelength Range 190 nm ̴ 800 nm Wavelength Repeatability ≤ 0.1 nm Wavelength Accuracy ≤± 1.0 nm Band Width 6 nm Response Time 0.0 s ̴ 5 s Baseline Noise ≤ ± 0.25 × 10 -5AU (empty cell, 254 nm, 1.0 s) Drift ≤0.5 × 10-4AU/h (empty cell, 254nm, 60min) Path length 10 mm(analytical cell) Light Source Deuterium Lamp and Tungsten Lamp Power Supply AC110/220V±10%, 50/60Hz |
High performance liquid chromatography MHPLC-3B
| Pump specifications | Type Gradient pump Flow Rate Range 0.001 ̴ 10.000 ml / min, in 0.001 mL increments Flow Accuracy ≤ ± 0.2 % (1.00 ml / min, H2O, R.T., 8.5 MPa) Flow Precision RSD ≤ 0.06 % (1.00 ml / min, H2O, R.T., 8.5 MPa) Maximum Pressure 63 MPa Pressure accuracy ≤ ± 3.0 % Pressure pulsation ≤ 0.1 MPa Composition rang 0 % ̴ 100 % Composition precision ≤ 0.2 % RSD |
| Detector specifications | Wavelength range 190 nm ̴ 800 nm Wavelength repeatability ≤ 0.1 nm Wavelength accuracy ≤ ± 1.0 nm Band width 6 nm Response time 0.0 sec ̴ 5 sec Baseline noise ≤ ± 0.25 × 10-5 AU (empty cell, 254 nm, 1.0 s) Drift ≤0.5 × 10-4 A U / h (empty cell, 254 nm, 60 min) Path length 10 mm (analytical cell) Light source Deuterium Lamp and Tungsten Lamp Power supply AC 110 V / 220 V ± 10 %, 50 Hz / 60 Hz |
High performance liquid chromatography MHPLC-3C
| Pump specifications | Type Quaternary low pressure pump Flow rate range 0.001 ̴ 10.000 mL / min, in 0.001 mL increments Flow accuracy ≤ ± 0.2 % (1.00 ml / min, H2O, R.T., 8.5 MPa) Flow precision RSD ≤ 0.06 % (1.00 ml / min, H2O, R.T., 8.5 MPa) Maximum pressure 63 MPa Pressure accuracy ≤ ± 3.0 % Pressure pulsation ≤ 0.1 MPa Composition rang 0 ̴ 100 % Composition precision ≤ 0.3 % RSD |
| Detector specifications | Wavelength range 190 nm ̴ 800 nm Wavelength repeatability ≤ 0.1 nm Wavelength accuracy ≤ ± 1.0 nm Band width 6 nm Response time 0.0 sec ̴ 5 sec Baseline noise ≤ ± 0.25 × 10-5 AU (empty cell, 254 nm, 1.0 s) Drift ≤ 0.5 ×10-4 AU / h (empty cell, 254 nm, 60 min) Path length 10 mm (analytical cell) Light source Deuterium Lamp and Tungsten Lamp Power supply AC 110 V / 220 V ± 10%, 50 Hz / 60 Hz |